Preview

General Reanimatology

Advanced search

Postoperative Prevention of Nosocomial Pneumonias, by Using Airway Filters under Artificial Ventilation

https://doi.org/10.15360/1813-9779-2007-3-95

Abstract

Objective: to determine the duration of exposition of filters and their most optimum position in different types of respiratory circuits. Subjects and methods. Group 1 comprised patients receiving not more than 12-hour artificial ventilation (AV) through a coaxial respiratory-circuit, without a humidifier, moisture accumulators, and a nebulizer. The filters were set between the intubation tube and the circuit. Group 2 included patients having more than 12-hour AV through the disposal respiratory circuits including a nebulizer, a humidifier, and inspiratory and expiratory moisture accumulators. The filters were set in front of the inspiratory circuit and behind the expiratory one on an AV apparatuses. Before regularly replacing a filter and/or a circuit, samples were taken for bacteriological tests from different portions of the circuit at various intervals: 4, 12, 24, 48, 96, 120, and 144 hours after initiation of AV, as well as before tracheal extubation in a patient. Results. The data obtained during the study have indicated that the breathing circuit filters afford a reliable protection from the entry of environmental microorganisms into the patient’s respiratory tract. In Group 1, AV did not result in the occurrence of pneumonia or tracheobronchitis. The moisture accumulators and the Y-shaped connector are the most infection-susceptible parts of a respiratory circuit. Mechanical circuit contamination and a larger number of pathogenic microbial strains were observed after 4—5-day AV. Conclusion. All procedures associated with circuit seal failure should be performed, by observing the aseptic rules. The recommended time of using a respiratory circuit during prolonged AV is 96—120 hours. Key words: intensive care unit, artificial ventilation, nosocomial pneumonia, antibacterial respiratory circuit filter.

References

1. Гельфанд Б. Р., Гологорский В. А., Белоцерковский Б. З. и др.

2. Craven D. E., Steger K. A., La Force F. M.Pneumonia. In: Bennett L. V., Brachman P. S. (eds), Hospital infections, 4th ed. Philadelphia: Lippincott-Raven press; 1998.

3. FagonJ.-Y., ChastreJ.Ventilator-associated pneumonia. Am. J. Respir. Crit. Care Med. 2002; 165: 867—90

4. Vincent J.-L.Microbial resistance: lessons from EPIC study. Intens. Care Med. 2000; 26 (1): 3—8.

5. Fagan J. Y., Chastre J., Domart Y. et al.Nosocomial pneumonia in patients receiving continuous mechanical ventilation. Am. Rev. Respir. Dis. 1989; 139: 877.

6. Гельфанд Б. Р., Белоцерковский Б. З., Гельфанд Е. Б. и др.Лечение но-зокомиальной пневмонии, связанной с искусственной вентиляцией легких у хирургических больных. Consilium medicum 2000; 2 (1).

7. Еременко А. А., Зюляева Т. П., Божьева Л. В. и др.Принципы профилактики пневмонии, связанной с использованием искусственной вентиляции легких в отделениях реанимации и интенсивной терапии. Анестезиология и реаниматология 2001; 2: 16—19.

8. Estes R. J., Meduri G. U.Intens. Care Med. 1995; 21: 365—383.

9. Craven D. E., Connolly M. G., Lichtenberg D. A. et al.Contamination of mechanical ventilators with tubing changes every 24 or 48 hours. N. Engl. J. Med. 1982; 306: 1505—150

10. Fagan J. Y., Chastre J., Hance A.J. et al.Evaluation of clinical judgment in the identification and treatment of nosocomial pneumonia in ventilated patients. Chest 1993; 103: 547—553.

11. Marquette C. H., Copin M. C., Wallet F. et al.Diagnostic tests for pneumonia in ventilated patients: prospective evaluation of diagnostic accuracy using histology as a diagnostic gold standard. Am. J. Respir. Crit. Care Med. 1995; 151: 1878—1888.

12. RelloJ., Ausina V., Ricart M. et al.Impact of previous antimicrobial therapy on the etiolog and outcome of ventilator-associated pneumonia. Chest 1993; 104: 1230—1235.

13. ChastreJ., Viau F., Brun P. et al.Prospective evaluation of the protected specimen brush for the diagnosis of pulmonary infections in ventilated patients. Am. Rev. Respir. Dis. 1984; 130: 924—929.

14. Craven D. E., Kunches L. M., Lichtenberg D. A. et al.Nosocomial infection and fatality in medical and surgical intensive care unit patients. Arch. Intern. Med. 1988; 148: 1161.

15. Roustan J. P., KienlenJ., Aubas P. et al.Comparision of hydrophobic heat and moisture exchangers with heated humifiders during prolonged mechanical ventilation. Intens. Care Med. 1992; 18: 97—100.

16. Salata R. A., Lederman M. M., Shlaes D. M. et al.Diagnosis of nose pneumonia in intubated, intensive care unit patients. Am. Rev. Dis. 1987; 135: 426.

17. Seidenfeld J. J., Pohl D. F., Bell R. C. et al.Incidence, site, and outcom infections in patients with the adult respiratory distress syndrome. Rev. Respir. Dis. 1986; 134: 12.

18. Fabregas N., Torres A., El-Ebiary M. et al.Histopathologic and microbi-ologic aspects of ventilator-associated pneumonia. Anesthesiology 1996; 84: 760—771.

19. Cook D., deJonghe B., Brochard L., Brun-Buisson C.Influence of airway management on ventilator-associated pneumonia. JAMA 1998; 279: 781—787.


Review

For citations:


Yeremenko A.A., Zorin D.Ye., Yegorov V.M., Kolpakov P.Ye. Postoperative Prevention of Nosocomial Pneumonias, by Using Airway Filters under Artificial Ventilation . General Reanimatology. 2007;3(3):95. https://doi.org/10.15360/1813-9779-2007-3-95

Views: 4425


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1813-9779 (Print)
ISSN 2411-7110 (Online)